Anechoic Chamber Suite
The Anechoic Chamber Suite, consisting of a radio frequency-shielded, EM absorber-lined chamber (10.3 x 6.8 x 3.7 m) and measurement equipment, enables simultaneous exploration of radiation, near-field, and far-field interactions of antennas and objects under test. This unique facility covers a wide frequency range from 900 MHz to 500 GHz.

Modular software and a PC control the Vector Network Analyzer (VNA), translational, and rotational stages, automating the measurement acquisition process. The software allows for the reconstruction of antenna radiation patterns from far-field (FF) and spherical near-field (SNF) measurements and supports the exploration of 3D monostatic and bistatic scattering profiles. Additionally, the software is fully customizable and scriptable to support non-standard measurements.
A VNA with a large dynamic range is used to coherently generate and detect RF signals from 900 MHz to 50 GHz. For frequencies beyond 50 GHz, frequency extender heads (add-on modules) are used to extend the VNA’s operational frequency range into the upper mm-wave bands, up to 500 GHz.
Key equipment:

- Rohde&Schwartz ZNA50 4 port Vector Network Analyzer (10 MHz – 50 GHz).
- Rohde&Schwartz and VDI Frequency Extension Modules (50 GHz – 500 GHz).
- Radio frequency shielded, and EM absorber lined chamber (10.3 x 6.8 x 3.7 m).
- Three platforms, each equipped with a stack of high-precision electromechanical stages, provide rotational and translational degrees of freedom for mounting antennas, RCS targets, or (meta-)material samples. Two of these platforms are moveable: one on a semi circular rail to support a feed for radar cross-section (RCS) measurements and the second on a linear rail to allow reconfiguration between far-field, RCS, and near-field measurement modes, providing either another feed location or support for a target or material.
Frequency Range/Dynamic Range
|
Frequency, GHz |
1 |
2 |
4 |
8 |
12 |
18 |
26 |
40 |
50 |
75 |
110 |
140 |
220 |
330 |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Dynamic Range, dB |
94 |
84 |
82 |
72 |
72 |
90 |
71 |
88 |
90 |
86 |
83 |
80 |
76 |
63 |
Our services

At the Anechoic Chamber suite, we offer a flexible range of services for industrial and academic research and development projects. Our equipment enables the exploration of simultaneous radiation, near-field, and far-field interactions of antennas and nearby objects, covering a frequency range from 900 MHz to 500 GHz.
It is uniquely equipped for electromagnetic (EM) characterization of antennas and novel EM structures (e.g., metasurfaces) and EM scattering in a controlled, reconfigurable environment.
The chamber, mechanically isolated from the building with vibrationally damped foundations and lined with absorber panels rated up to 500 GHz, provides seamless hardware-software integration for automated, high-precision measurements with a large dynamic range.
We also provide technical support for system usage and software tools.
| Band | L | L, S | S | C | C | X | Ku | K | Ka | V | V | W | mm | mm | mm |
| Frequency min, GHz | 1.1 | 1.7 | 2.6 | 3.95 | 5.85 | 8.2 | 12.4 | 18 | 26.5 | 40 | 50 | 75 | 110 | 140 | 220 |
| Frequency max, GHz | 1.7 | 2.6 | 3.95 | 5.85 | 8.2 | 12.4 | 18 | 26.5 | 40 | 60 | 75 | 110 | 170 | 220 | 325 |
| Antenna gain, dBi | 15 | 20 | 20 | 20 | 20 | 23 | 23 | 23 | 23 | 23 | 25 | 25 | 25 | 25 | 25 |
| Flange / connector | SMA | SMA | SMA | SMA | SMA | SMA | SMA | 3.5mm | 2.92mm | 1.85mm | WR15 | WR10 | WR6 | WR5 | WR3 |
Frequency Bands and Antennas
For each frequency band two standard gain horn antennas and one waveguide probe/conical horn are available.